How to use 12d hifu machine

Warning

Dear user:

Thanks for using our company's products, in order to fully make use of the products,

we sincerely suggest the following:

Please read the instructions in detail and keep it well for further reading and consulting.

Please follow the instructions to fit and operate the machine correctly.

Please do not remove or change any accessories of the machine.

It forbids others to open and remove the mainframe except the authorized people in the company.

It forbids others to change and maintain the machine except professional people.

Please contact us if there is any trouble during the operation


This product does not belong to medical,
please don’t use for disease treatment!


Parts Introduction

Accessories

Attention: the Work head (consumable) is rated 20000 times, When the times is less than 1000, no longer provide a work head maintenance service, please replace the new work head.

Any part of the instrument made of PMMA material (for example: accessories shelf, panel, transparent handles, etc..). Do not use alcohol or corrosive solvent to wipe clean.


Basal Operation 

1.Put the accessories well and connect them to the host:

(1)Insert the SD handle into [3]jack.

(2)Insert the LIPO handle into [4]jack.

(3)Insert the FL handle into [5]jack.

(4)Insert the foot switch into [6]jack.

(5)Insert the DEP handle into [5]jack.

.2.Insert the plug of Power supply line into [8] socket, before the first use, first, install the work

handle, turn on Power switch [9]. Pay attention to turn on/off the system must interval 30 seconds

each time to ensure the normal use of the system.

3.Scan the code at startup and activate the password according to the prompt. After activation, the

password is "000000" by default, enter password,the system enter interface as (Fig 1) shows.

723496695ce9f89f449290b633bfe8e

4.In the interface as (Fig 1):

(1)Click on “Basic setting” to select language, set time and date. Click “EXIT” to return (Fig 1).

Attention: “Advanced setting” is only for engineering debugging, not for user.

(2)Click on “FL ” icon, the system enter FL operating interface as (Fig 2) shows.

(3)Click on “ SD” icon, the system enter SD operating interface as (Fig 4) shows.

(4)Click on “LIPO ” icon, the system enter LIPO operating interface as (Fig 7) shows.

(5)Click on “DEP ” icon, the system enter DEP operating interface as (Fig 8) shows.



FL OPERATE

1. Select a suitable working head and insert the direction of the handle shear head into the slot of the handle. On the contrary, if the working head needs to be replaced, press the handle unlocking button [9] and then pull out the working head for replacement. The working head can be automatically identified.

80e6b7b99088bd97d94b8926309ec20

2. On the interface as shown in Fig 2:

(1) The "WD6" position displays the current reading information of the working head. After replacing the working head, the system will automatically recognize the model of the working head.

(2) The first data represents the number of transmissions used by the working head. The second data represents the number of calibration launches of the working head.

(3) "Current emission count" represents the number of emissions ofthe current course operation. Click "clear " to clear the data of "current

transmission times".

(4) The operation of this instrument is divided into "arm", "abdomen", "waist" and "thigh". Select "arm" to enter the divided area on the

schematic diagram of the operation interface (Fig. 2) for block nursing operation. The data displayed in each area on the interface is the number of times that the area currently receives the work head. Click"clear “, all data in each area can be cleared.

12D HIFU(Fig 2)

(5) Select mode: automatic mode or manual mode. Click "automatic" to enter automatic mode

(Fig2):

Set power: the default output power of the working head is 1.0j, ranging from 0.1j (weak) to 3.0j (strong). Click "<" and ">" to set.

Set point distance: range: 1.0-5.0, each gear point distance is 0.5, click "<" and ">" to set, and the default is 1.5.

Set the number of lines: range: 2-12 lines. Click "<" and ">"to set. The default is 6 lines.

Set length: range: 5-25mm, default is 25, click "<" and ">" to set.

Setting total number of rows: range: 1-100, click "<" and ">"to set.

Click "manual" to enter manual mode (Fig 3):

Fig 3

Set the power. The default output power of the working head is 1.0j, ranging from 0.1j (weak) to 3.0j (strong). Click "<"

and ">" to set.

Set point distance: range: 1.0-5.0, each gear point distance is 0.5, click "<" and ">" to set, and the default is 1.5.

Set the number of lines: range: 2-12 lines. Click "<" and ">" to set. The default is 6 lines.

Set length: range: 5-25mm, default is 25, click "<" and ">" to set.


3. The operator clicks the "on" button, and the system enters the state to be launched. At this time,

the indicator light on the fl handle is normally on.

4. Click a certain area, and then press the start button on the fl handle once (or step on the pedal on

/ off device), the working head will emit a line of points and record the emission count in this area

5. In the state to be launched, the length is not adjustable.

6. Click the "pause" button to enter the pause state. At this time, press the launch button (or pedal

on / off) on the fl handle, and the system will not work

7. Click "home " to return to fig 1.

Operation suggestions: transmitter working heads S1.5,S2.0,S4.5,WD3.0 are used for face

operation, and transmitter working heads WD4.5,WD6,WD9,WD12 are used for body

operation.The face operation method is the same as the body operation. Note that the " " part in

the figure below is the prohibited operation area when the face operation is performed

the prohibited operation area when the face operation is performedFig 5

Cautions after Facelifting

1.Strengthen the moisturizing and sunscreen.

2.DO NOT take any medicine and drink alcohol before and after 24 hours. Follow the nursing advice.

3.DO NOT soak hot springs and have a steam bath.

4.It is normal that the skin feel dry after treatment. Pay attention to water replenishing. Suggestions for put

replenishment mask at least three times a week



Adapt to The Crowd

1.people who have facial skin sagging, wrinkles, skin aging and rough, reshape the face curve, Effectively

improve the eye region and neck tissue sag, ptosis.

2.Tightening the loose skin, removing real and false wrinkles, repair stretch marks.

3.Acnes.

4.Suitable for a variety of wrinkles and deep wrinkles, forehead wrinkles, eye wrinkles, purify end, nasolabial

fold, lip wrinkles, frown lines etc.


Cautions


1. Before nursing, fill in a customer inquiry form and ask the customer to sign for confirmation.

Take photos before operation for comparison after nursing.

2. Draw the area to be operated on the face or neck with a makeup pen. Avoid the forbidden area and do not operate in the forbidden area.

3. Put the matrix ruler on the abdomen and draw the operable parts with a makeup pen.

4. Clean the skin before operation to make it moist. Apply water replenishing facial mask after operation.

5. The principle of selecting a suitable working head is to select it according to the operating part.

When replacing the handle working head, the power must be turned off first.

6. Apply sufficient gel on the operation part as the coupling agent between the working head and the skin.

7. Click the start button "start" on the operation interface, and the system is in the state of waiting for launch. During operation, the working head must be vertically close to the skin. Press the launch button (or step on the foot switch) on the fl handle. Each time the launch button is pressed, a line of points will be formed at the specified depth under the cortex. During operation, attention must be paid to ensure that there is enough gel between the working head and the skin. During the operation, the area is divided according to the part and the

emission is completed line by line until it is moved to the specified position.

Note " !" is a special note.

Clean the used working head after nursing. Cleaning method: dip soft objects such as cotton swabs and paper towels in clean water and gently wipe the black conductive film from the middle part to both sides. Do not use strong force to avoid damaging the black conductive

film. Do not use alcohol-containing products and disinfectants to clean, so as to prevent the black conductive film from falling. Do not tear the black conductive film.

NOTE:

1. It is forbidden to operate without coupling agent or gel.

2. The working head cannot be launched without load to avoid damaging the working head accessories. Each launch shall ensure that the black conductive film of the working head is coated with uniform medium.

3. Wipe the essential oil liquid on the handle in time after use. Do not let the liquid penetrate into the inside of the working handle to prevent the ha